%0 Journal Article %T Comparison of the 3D triple-scan protocol and the impression replica technique for 3-unit tooth-supported fixed dental prostheses %A Mir %A Per Svanborg %A Therese Reinedahl %A Torgny Alstad %A a Andersson %J Biomaterial Investigations in Dentistry %D 2019 %R https://doi.org/10.1080/26415275.2019.1684198 %X Abstract Objective: The aim of this study was to compare the fit measurements from the 3£żD triple-scan protocol and the impression replica technique, by measuring the fit of three-unit tooth-supported fixed dental prostheses. Materials and methods: The results from an earlier study using the triple-scan protocol were compared to results from the impression replica technique for absolute marginal gap, cervical area gap and internal gap. Results: There were no differences for cervical area gap or internal gap, but for absolute marginal gap the impression replica technique obtained significantly smaller fit values. Conclusion: The impression replica technique and the triple-scan protocol may both be used to measure the fit of tooth-supported restorations. However, the scanner used for the triple-scan protocol must be able to obtain scan points at the outermost edge of the restorations when used for absolute marginal gap %U https://www.tandfonline.com/doi/full/10.1080/26415275.2019.1684198