%0 Journal Article %T Attenuation length measurements of liquid scintillator with LabVIEW and reliability evaluation of the device %A L. Gao %A BX. Yu %A YY. Ding %A L. Zhou %A LJ. Wen %A YG. Xie %A ZG. Wang %A X. Cai %A XL. Sun %A J. Fang %A Z. Xue %A AW. Zhang %A QW. L¨¹ %A LJ. Sun %A YS. Ge %A YB. Liu %A SL. Niu %A T. Hu %A J. Cao %A JG. L¨¹ %J Physics %D 2013 %I arXiv %R 10.1088/1674-1137/37/7/076001 %X The attenuation length measuring device was constructed by using oscilloscope and LabVIEW for signal acquisition and processing. The performance of the device has been tested with a variety of ways, the test results show that the set-up has a good stability and high precision (sigma/mean reached 0.4 percent). Besides, the accuracy of the measurement system will decrease by about 17 percent if a filter is used. The attenuation length of gadolinium-loaded liquid scintillator (Gd-LS) was measured as 15.10 plus or minus 0.35 m where Gd-LS was heavily used in Daya Bay Neutrino Experiment. In addition, one method based on the Beer-Lambert law was proposed to investigate the reliability of the measurement device, the R-square reached 0.9995. Moreover, three purification methods for Linear Alkyl Benzene (LAB) production were compared in the experiment. %U http://arxiv.org/abs/1305.1471v1