%0 Journal Article %T Dose Validation of Physical Wedged Asymmetric Fields in Artiste Linear Accelerator %A Tamer Dawod %A E. M. Abdelrazek %A Mostafa Elnaggar %A Rehab Omar %J International Journal of Medical Physics,Clinical Engineering and Radiation Oncology %P 201-209 %@ 2168-5444 %D 2014 %I Scientific Research Publishing %R 10.4236/ijmpcero.2014.34026 %X Aim: The purpose of this study was to make a comparison between measured and calculated physical wedge dose distributions using the superposition algorithm. Settings and Design: The accurate determination of absorbed dose is important radiotherapy because of the relatively steep sigmoidal dose response curves for both tumor control and normal-tissue damage. Materials and Methods: High-energy photons (6 and 10 MV) from Artiste Treatment System Linear Accelerator Machine, available at Alexandria Ayadi Al-Mostakbal Oncology Center, were used. Results and Discussion: The results showed that the difference between measured and calculated wedged isodose curves depends on field size, beam energy, and the angle of the used wedge. Conclusion: The results showed that the presence of a wedge alters the primary and scattered components generated by a linear accelerator and causes beam hardening in 6 and 10 MV. The beam hardening increased as the wedge angle increased. %K Radiotherapy %K Wedge %K Linear Accelerator %K Computerized Treatment Planning System %U http://www.scirp.org/journal/PaperInformation.aspx?PaperID=50546