%0 Journal Article %T Accuracy of Medical Oxygen Flowmeters: A Multicentric Field Study %A Fr¨¦d¨¦ric Duprez %A Maria Barile %A Thierry Bonus %A Gr¨¦gory Cuvelier %A Sandra Ollieuz %A Shahram Mashayekhi %A Alexandre Legrand %J Health %P 1978-1983 %@ 1949-5005 %D 2014 %I Scientific Research Publishing %R 10.4236/health.2014.615232 %X The accuracy of 476 oxygen flowmeters was investigated using a thermal mass flowmeter in eight hospitals in France and Belgium. Different oxygen flow rates (2 to 15 l/min) were evaluated at the patient¡¯s bed. When the sample was considered as a whole, the accuracy of delivered flow was acceptable but precision was poor. The variability of the delivered flow between devices was greater when a low flow rate was required. Compensated-pressure oxygen flowmeters for these low rates were more accurate than their non-compensated counterparts. This study emphasizes the need to individually adapt the oxygen flow rate each time a patient has to move from one flowmeter to another. %K Medical Device Accuracy %K Oxygen Therapy %K Flowmeter %U http://www.scirp.org/journal/PaperInformation.aspx?PaperID=48743