%0 Journal Article %T Is It Possible to Publish a Calibration Function for Radiochromic Film? %A Maria F. Chan %A David Lewis %A Xiang Yu %J International Journal of Medical Physics,Clinical Engineering and Radiation Oncology %P 25-30 %@ 2168-5444 %D 2014 %I Scientific Research Publishing %R 10.4236/ijmpcero.2014.31005 %X
Purpose: To assess the possibility of using a public calibration function for radiochromic film dosimetry in dose QA of highly conformal treatment plans. Methods: EBT3 film calibration strips (3.5 กม 20 cm2 from lots A101212 and A011713) were exposed on a Varian Trilogy at a facility to a 10 กม 10 cm2 open field at doses of 80, 160, 320 cGy using 6MV photons. Together with a strip of unexposed film from the same lot, the exposed films were digitized in a single scan using different Epson 10,000 XL scanners at two different facilities. The dose-response data for each color-channel from each facility were generated using the same calibration function X(D) = a + b/(D - c), where X(D) is the response at dose D and a, b and c are the coefficients. Different batches of EBT3 film were exposed to a VMAT beam. These films, plus two reference strips exposed to doses of zero and 160 cGy, were digitized on the scanners at the two facilities. Using the multi-channel dosimetry method and One-scan protocol %K Dosimetry %K Quality Assurance %K IMRT %K Radiochromic Film %K Multichannel %U http://www.scirp.org/journal/PaperInformation.aspx?PaperID=42613