%0 Journal Article %T A newly designed big cup nitinol stent for gastric outlet obstruction %A Ding Shi %A Sheng-Hui Liao %A Jian-Ping Geng %J World Journal of Gastroenterology %D 2010 %I Baishideng Publishing Group Co. Limited %X AIM: To find out whether a newly designed big cup nitinol stent is suitable for treatment of patients with gastric outlet obstruction resulting from gastric cancer.METHODS: The new stent is composed of a proximal big cup segment (20 mm in length and 48-55 mm in diameter), a middle part (60 mm in length and 20 mm in diameter) covered by a polyethylene membrane and a distal sphericity (20 mm in length and 28 mm in diameter). Half of the proximal big cup segment is also covered by a polyethlene membrane, which is adjacent to the middle part of the stent. The stent is preloaded in a 6.0-mm-diameter introducer system. Thirteen patients with gastric outlet obstruction resulting from gastric cancer received the new stents under endoscopic and fluoroscopic guidance.RESULTS: Technical success was achieved in 12 of 13 (92.3%) patients. Among the 12 patients in whom endoscopic stent was placed successfully, the clinical success rate was 91.7% during a follow-up of average 6.5 mo. During the first month follow-up, the migration rate was 0%, recurrent obstruction 0% and gastric bleeding 8.3%. During the follow-up between 2-12 mo, no migration, recurrent obstruction and gastric bleeding occurred.CONCLUSION: The proximal big cup segment seems to be effective and promising for technical efficacy, clinical outcome, and preventing migration and tumor ingrowth and increasing the emptying rate of sinus ventriculi. %K Endoscopic %K Gastric outlet %K Stenosis %K Obstruction %K Stents %U http://www.wjgnet.com/1007-9327/full/v16/i33/4206.htm