%0 Journal Article
%T Determination of a-Linolenic Acid in Perilla Oil by Reversed- Phase High Performance Liquid Chromatography Coupled with Evaporative Light-Scattering Detector
反相高效液相色谱-蒸发光散射检测法测定紫苏子油中的α-亚麻酸
%A WANG Qiao'e
%A SONG Ping
%A DING Mingyu
%A
王巧娥
%A 宋萍
%A 丁明玉
%J 色谱
%D 2006
%I
%X A method for the determination of alpha-linolenic acid in perilla oil was developed using the reversed-phase high performance liquid chromatography coupled with evaporative light-scattering detector (RP-HPLC-ELSD). The perilla oil was saponified by 0.5 mol/L KOH-CH3OH solution for 20 min in a 60 degrees C water bath, then acidified by 6 mol/L HCl and finally the dissociative fatty acids, including alpha-linolenic acid, was extracted by anhydrous ether. After the ether was blown out by nitrogen, the residuals were dissolved by 10 mL methanol. The calibration curve was found to be linear over the range of 6.2 - 45.4 microg (r = 0.997 3, n = 5) and the detection limit was 0.11 microg (S/N = 3). The average recovery was 102% and their relative standard deviation (RSD) was 6.3% (n = 5). The content of alpha-linolenic acid in the determined perilla oil was 6.79% which is consistent with the previous report.
%K 反相高效液相色谱
%K 蒸发光散射检测器
%K α-亚麻酸
%K 紫苏子油
%U http://www.alljournals.cn/get_abstract_url.aspx?pcid=6E709DC38FA1D09A4B578DD0906875B5B44D4D294832BB8E&cid=6579068328FE643F&jid=4D81E042D77AFEC6881D14759692069C&aid=B2311C01229D71F9&yid=37904DC365DD7266&vid=B91E8C6D6FE990DB&iid=CA4FD0336C81A37A&sid=E514EE58E0E50ECF&eid=11B4E5CC8CDD3201&journal_id=1000-8713&journal_name=色谱&referenced_num=0&reference_num=14