%0 Journal Article %T Determination of aromatic hydrocarbon types in diesel oil using high performance liquid chromatography with photodiode array detection
高效液相色谱-光电二极管阵列检测法测定柴油中芳烃的类型 %A LIN Yu %A LI Yi %A BAI Zhengwei %A HE Chengyue %A ZHANG Qi %A
林玉 %A 李怿 %A 白正伟 %A 何成跃 %A 张奇 %J 色谱 %D 2008 %I %X When using ASTM D6591 to determine the content of aromatic hydrocarbons in diesel oil, a photodiode array detector (PDAD) was used to monitor the separation of aromatic hydrocarbons. The influence of back flush time B was studied and it was observed the results of di-aromatic hydrocarbons or poly-aromatic hydrocarbons will give errors when the back flush time is more than B +/- 0.2 min. The PDAD can also be used to determine the best cutting time between mono-aromatic hydrocarbons and di-aromatic hydrocarbons, when the separation was not baseline resolved. %K high performance liquid chromatography (HPLC) %K photodiode array detection %K aromatic hydrocarbon types %K diesel oil
高效液相色谱 %K 光电二极管阵列检测 %K 芳烃类型 %K 柴油 %K 高效液相色谱 %K 光电二极管阵列检测器 %K 二极管阵列检测法 %K 测定 %K 柴油 %K 双环芳烃 %K 类型 %K detection %K photodiode %K array %K liquid %K chromatography %K high %K performance %K diesel %K oil %K aromatic %K hydrocarbon %K 切割点 %K 单环芳烃 %K 范围 %K 变化 %K 发现 %K 影响 %K 结果 %U http://www.alljournals.cn/get_abstract_url.aspx?pcid=6E709DC38FA1D09A4B578DD0906875B5B44D4D294832BB8E&cid=6579068328FE643F&jid=4D81E042D77AFEC6881D14759692069C&aid=B0465FCD86DF11EFC6F69A9B0331BB32&yid=67289AFF6305E306&vid=96C778EE049EE47D&iid=0B39A22176CE99FB&sid=1D01216AD76577EC&eid=82722E2B785EBF0D&journal_id=1000-8713&journal_name=色谱&referenced_num=0&reference_num=12