%0 Journal Article %T Simultaneous Chronoamperometric Sensing of Ascorbic Acid and Acetaminophen at a Boron-Doped Diamond Electrode %A Codru£ża Cofan %A Ciprian Radovan %J Sensors %D 2008 %I MDPI AG %R 10.3390/s8063952 %X Cyclic voltammetry (CV) and chronoamperometry (CA) have been used to sense and determine simultaneously L-ascorbic acid (AA) and acetaminophen (AC) at a boron-doped diamond electrode (BDDE) in a Britton-Robinson buffer solution. The calibration plots of anodic current peak versus concentration obtained from CV and CA data for both investigated compounds in single and di-component solutions over the concentration range 0.01 mM ¨C 0.1 mM proved to be linear, with very good correlation parameters. Sensitivity values and RSD of 2-3% were obtained for various situations, involving both individual and simultaneous presence of AA and AC. The chronoamperometric technique associated with standard addition in sequential one step and/or two successive and continuous chronoamperograms at two characteristic potential levels represented a feasible option for the simultaneous determination of AA and AC in real sample systems such as pharmaceutical formulations. The average values indicated by the supplier were confirmed to a very close approximation from chronoamperomgrams by using several additions with the application of suitable current correction factors. %K acetaminophen %K ascorbic acid %K chronoamperometry %K boron-doped diamond electrode %K buffered media. %U http://www.mdpi.com/1424-8220/8/6/3952