%0 Journal Article
%T Reflectivity uncertainty analysis of planar mirror calibration in BSRF
同步辐射标定平面镜反射率不确定度分析方法研究
%A Chen Bo-Lun
%A Yang Zheng-Hu
%A Cao Zhu-Rong
%A Dong Jian-Jun
%A Hou Li-Fei
%A Cui Yan-Li
%A Jiang Shao-En
%A Yi Rong-Qing
%A Li San-Wei
%A Liu Shen-Ye
%A Yang Jia-Min
%A
陈伯伦
%A 杨正华
%A 曹柱荣
%A 董建军
%A 侯立飞
%A 崔延莉
%A 江少恩
%A 易荣清
%A 李三伟
%A 刘慎业
%A 杨家敏
%J 物理学报
%D 2010
%I
%X A key diagnostic for absolute radiation flux and radiation temperature in hohlraum experiments is the Dante soft x-ray spectrometer in which grazing incidence x-ray mirrors are used to diminish the high energy response of the filters. The reflectivities of planar mirrors made of different materials were calibrated in BSRF in 2009.Considering both the influence of the source such as the spectra resolution,harmonic and the attenuation of the source intensity and the influence of the coherence of two detectors responses,we give the corrected mirror reflectivities and the uncertainty, including the statistical uncertainty of the expermental data.
%K synchrotron radiation
%K reflectivity of planar mirror
%K calibration
%K uncertain analysis
同步辐射
%K 平面镜反射率
%K 标定
%K 不确定度
%U http://www.alljournals.cn/get_abstract_url.aspx?pcid=6E709DC38FA1D09A4B578DD0906875B5B44D4D294832BB8E&cid=47EA7CFDDEBB28E0&jid=29DF2CB55EF687E7EFA80DFD4B978260&aid=77245ABE471EF4E9E4134DE3F589299A&yid=140ECF96957D60B2&vid=6AC2A205FBB0EF23&iid=F3090AE9B60B7ED1&sid=9B4CAAFB99240531&eid=7CEC80375CE02365&journal_id=1000-3290&journal_name=物理学报&referenced_num=0&reference_num=20