%T Assessment on the Accuracy of Four Types of Pulse by SM-1A TCM Pulse Detector
SM-1A中医四脉脉诊仪脉象诊断准确性的评价
%A Hu Jingqing
%A Zhao Ting
%A Xu Guicheng
%A Wu Lie
%A Kou Qiu''ai
%A Zhuang Hong
%A Li Zhenhu
%A Su Mingliang
%A Zhang Guangfu
%A
胡镜清
%A 赵婷
%A 徐贵成
%A 吴烈
%A 寇秋爱
%A 庄红
%A 李振华
%A 宿明良
%A 张广福
%J 世界科学技术-中医药现代化
%D 2011
%I
%X This study aimed to assess the reproducibility of SM-1A TCM pulse detector according to traditional Chinese medicine (TCM) experts. A total of 120 subjects were enrolled. Each subject was diagnosed by the SM-1A TCM pulse detector and TCM expert, respectively. And the pulse diagnosis was based on four types of pulse, including yimai,ju-mai, ren-mai and wen-mai. The Kappa test was applied. Compared to TCM expert, the sensitivity of the pulse detector for yi-mai was 0.859, for ju-mai was 0.945, for ren-mai was 0.957, and for wen-mai was 0.898. The specificity of the pulse detector for yi-mai was 0.875, for ju-mai was 0.8, for ren-mai was 0.907, and for wen-mai was 0.833. The Kappa value for yi-mai was 0.733, for ju-mai was 0.631, for ren-mai was 0.763, and for wen-mai was 0.549. The coincidence rate for yi-mai was 0.867, for ju-mai was 0.933, for ren-mai was 0.917, and for wen-mai was 0.892. The Youden index for yi-mai was 0.734, for ju-mai was 0.745, for ren-mai was 0.864, and for wen-mai was 0.731. The coincidence rate of SM-1A TCM pulse detector in the successive diagnosis of the same patient was 81.7%. Therefore, the SM-1A TCM pulse detector is able to diagnose the pulse accurately based on the four types of pulse. The test-retest reliability was good.
%K TCM pulse detector
%K Diagnostic test
脉诊仪
%K 诊断性试验
%U http://www.alljournals.cn/get_abstract_url.aspx?pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=B5F35A1FC27C1FF7&jid=44257A35316DB0727BA6BC707B9EDAAB&aid=656E32E02735604BFAC9E1B883181A4F&yid=9377ED8094509821&vid=FC0714F8D2EB605D&iid=CA4FD0336C81A37A&sid=67969BA850333433&eid=E44E40A2398D4F2A&journal_id=1674-3849&journal_name=世界科学技术-中医药现代化&referenced_num=0&reference_num=0