%T The Glutamate Biosensor and Its Application to Flow Injection Analysis System
谷氨酸传感器及在流动注射分析中的应用
%A Ye Bangce Li Qingshan Li Yourong Yu Juntang
%A
叶帮策
%A 李青山
%A 李友荣
%A 俞俊棠
%J 生物工程学报
%D 1994
%I
%X An enzyme electrode was prepared by cross-linking the L-glutamate oxidase with glutar-aldehyde on an aminopropyl-platinized-platinum wire. It was used for the determination of L-glutamate in a flow injection analysis system. The peak current is linearly related to the L-glutamate concentration in the range 0-2.0 mmol/L, with good performance: accuracy (CV = 0. 4%), fast response (<60sec), stability (> 20 days).
%K Biosensor
%K enzyme electrode
%K glutamate determination
%K glutamate oxidase
%K flow injection analysis
生物传感器,酶电极,谷氨酸测量,谷氨酸氧化酶,流动注射分析
%U http://www.alljournals.cn/get_abstract_url.aspx?pcid=90BA3D13E7F3BC869AC96FB3DA594E3FE34FBF7B8BC0E591&jid=A66E90C274451689E69F6F0291467824&aid=E73E28AFFFED27A16BE9A0C35B86B351&yid=3EBE383EEA0A6494&vid=F3090AE9B60B7ED1&iid=0B39A22176CE99FB&journal_id=1000-3061&journal_name=生物工程学报&referenced_num=1&reference_num=0