%0 Journal Article
%T Biosensor Based on Chemiluminescence for Serum Uric Acid Determination
尿酸酶-鲁米诺化学发光传感器
%A Liu Jianguo Guo Jian Li Gaoxiang
%A
刘建国
%A 郭健
%A 黎高翔
%J 生物工程学报
%D 1995
%I
%X A biosensor based on luminol chemiluminescent reaction and immobilized uricase column for serum uric acid determination has been developed. The response time by the sensor for various concentration of serum uric acid was 47 seconds with 17ul sample volume and at 40 samples per hour. The linear range of standard curve was from 1 up to 20 mg/dl of serum uric acid. The within day imprecision was 3.22%~4. 36%. The day to day imprecision was 6.18%-7.80%. The recovery rate by the method was 93% ~109%. Comparised with standard colorimetric method employed enzymatic kit revealed that the linear regression and correlation coerricient were Y=0. 4+0. 938X and r = 0. 9909 respectively. The immobilized uricase column retained 94% of its original activity even after over 2000 runs for five and half monthes' continual usage.
%K Uric acid sensor
%K chemiluminescence
%K flow injection analysis
%K immobilized enzyme
尿酸传感器,化学发光,流动注射分析,固定化酶
%U http://www.alljournals.cn/get_abstract_url.aspx?pcid=90BA3D13E7F3BC869AC96FB3DA594E3FE34FBF7B8BC0E591&jid=A66E90C274451689E69F6F0291467824&aid=773F4B74DCE6E587A3CF2CAE56161E32&yid=BBCD5003575B2B5F&vid=708DD6B15D2464E8&iid=38B194292C032A66&journal_id=1000-3061&journal_name=生物工程学报&referenced_num=0&reference_num=0