%0 Journal Article %T Comparison of CDE data in phacoemulsification between an open hospital-based ambulatory surgical center and a free-standing ambulatory surgical center %A Ming Chen %A Mindy Chen %J Clinical Ophthalmology %D 2010 %I %R http://dx.doi.org/10.2147/OPTH.S15076 %X mparison of CDE data in phacoemulsification between an open hospital-based ambulatory surgical center and a free-standing ambulatory surgical center Original Research (3802) Total Article Views Authors: Ming Chen, Mindy Chen Published Date November 2010 Volume 2010:4 Pages 1287 - 1289 DOI: http://dx.doi.org/10.2147/OPTH.S15076 Ming Chen1, Mindy Chen2 1University of Hawaii, Honolulu, HI, USA; 2University of California, Irvine, CA, USA Abstract: Mean CDE (cumulative dissipated energy) values were compared for an open hospital-based surgical center and a free-standing surgical center. The same model of phacoemulsifier (Alcon Infiniti Ozil) was used. Mean CDE values showed that surgeons (individual private practice) at the free-standing surgical center were more efficient than surgeons (individual private practice) at the open hospital-based surgical center (mean CDE at the hospital-based surgical center 18.96 seconds [SD = 12.51]; mean CDE at the free-standing surgical center 13.2 seconds [SD = 9.5]). CDE can be used to monitor the efficiency of a cataract surgeon and surgical center in phacoemulsification. The CDE value may be used by institutions as one of the indicators for quality control and audit in phacoemulsification. %K CDE (cumulative dissipated energy) %K open hospital-based ambulatory surgical center %K free-standing surgical center %K phacoemulsification %U https://www.dovepress.com/comparison-of-cde-data-in-phacoemulsification-between-an-open-hospital-peer-reviewed-article-OPTH