%0 Journal Article %T Optimization of Measuring Current for Chopper Low Resistance Comparator %A R. M. Radetic %A D. R. Milivojevic %A V. M. Despotovic %J Measurement Science Review %@ 1335-8871 %D 2010 %I %R 10.2478/v10048-010-0005-7 %X The article describes systematic error analysis for realized chopper low-resistance comparator. The objective was to determine optimal working conditions, considering limited resolution (about 10 nV) and resistance variation due to self-heating caused by increase of measuring current. The analysis and performances of realized chopper low-resistance comparator prototype instrument confirmed the initial assumptions of applied principles. The error for 1 A measuring current is 1.15 ppm in the measuring range of 10 m¦¸. It is possible to use the same principle for calculation of optimal measuring current and measurement error in other measurement ranges, as well. %K Low-resistance measurement %K chopper comparator %K error analysis %U http://versita.metapress.com/content/336n40j475p8v55n/?p=0fed901c71d24eccacbab7e094928001&pi=4