%0 Journal Article %T Rapigluco-test. Resultados de la evaluaci¨®n externa y pruebas de terreno %A Russeaux Gu¨ªa %A Niurka %A Vald¨¦s Diez %A Lilliam %A Bilbao Abraham %A Yuria %J Revista Cubana de Farmacia %D 2000 %I ECIMED %X the rapigluco-test was evaluated at the national center of reference for clinical laboratory. the lineality interval was successful between 2.5 and 22 mmol/l of glucose and the recovery of the method was of 90%. during the accuracy test it was found that both the experimental lots and the scale-up lots had a good correlation with the reference methods used. the precision evaluated at 2 levels of concentration was also good. 49 lots of the reagent were distributed in hospitals and polyclinics all over the country. 1 818 glucose determinations were obtained from 1993 to 1996. there were no complaints about the behaviour of the product. %K blood glucose %K indicators and reagents [analysis]. %U http://scielo.sld.cu/scielo.php?script=sci_abstract&pid=S0034-75152000000300001&lng=en&nrm=iso&tlng=en