%0 Journal Article %T Elabora£¿£¿o e implementa£¿£¿o de um programa para verifica£¿£¿o dos c¨¢lculos das unidades monitoras em radioterapia %A M¨¹ller %A M¨¢rcio Rog¨¦rio %A Rodrigues %A Laura Natal %A Silva %A Marco Ant£¿nio da %J Radiologia Brasileira %D 2005 %I Col¨¦gio Brasileiro de Radiologia %R 10.1590/S0100-39842005000600007 %X objective: to develop and implement a software based on the manual calculation algorithm for verification of the calculations in radiotherapy monitor units and to establish acceptance levels as a mechanism of quality assurance. materials and methods: data were obtained from clinac 600c and 2100c varian linear accelerators and the computerized treatment planning system used was cadplan£¿. results: for the 6 mv beams the acceptance levels for deviations among the calculations of monitor units, divided by treatment area, were the following: breast (0.0% ¡À 1.7%), head and neck (1.5% ¡À 0.5%), hypophysis (-1.7% ¡À 0.5%), pelvis (2.1% ¡À 2.1%) and thorax (0.2% ¡À 1.3%). for the 15 mv beams the suggested level for pelvis in all the treatment techniques was (3.2% ¡À 1.3%). conclusion: the present data are sufficient to justify the use of the software in the clinical practice as a tool for the quality assurance program. %K monitor units calculation %K quality assurance %K computerized treatment planning system. %U http://www.scielo.br/scielo.php?script=sci_abstract&pid=S0100-39842005000600007&lng=en&nrm=iso&tlng=en