%0 Journal Article %T Simultaneous Echo Refocused Magnetization Transfer Imaging %A Claus Kiefer %J Open Journal of Medical Imaging %P 154-158 %@ 2164-2796 %D 2014 %I Scientific Research Publishing %R 10.4236/ojmi.2014.43021 %X Purpose: To increase the efficiency of densely encoded magnetization transfer imaging of the brain, we time-multiplex multiple slices within the same readout using simultaneous echo refocusing FLASH imaging with magnetization transfer (MT) preparation (MT-SER-FLASH). Materials and Methods: Inefficiency in total scan time results from the number of frequency samples needed for sufficient quality of quantitative parameter maps for a binary spin bath model. We present a highly efficient multiplexing method, simultaneous echo refocused magnetization transfer imaging (MT-SER-FLASH) for reducing the total scan time of MT imaging by one-third. The specific absorption rate (SAR) was also reduced by reducing the number of MT-pulses per volume. Results: 2D-MT-SER-FLASH is performed in 19 minutes rather than 1 hour, acceptable for routine clinical application. The SAR could be reduced to 69% instead of more than 100% with a standard 2D or 3D-FLASH with MT-preparation. Conclusion: The net reduction of scan time and SAR enables the use of quantitative model based magnetization transfer imaging within a clinical environment. %K MRI %K Magnetization Transfer Imaging %K Fast Imaging %K Pulse Sequences %K Simultaneous Echo Refocusing %U http://www.scirp.org/journal/PaperInformation.aspx?PaperID=49917