%0 Journal Article %T Long Term Follow up of the Solyx Single Incision Sling in the Treatment of Female Stress Urinary Incontinence (SUI) %A Scott Serels %A Michael Douso %J Open Journal of Urology %P 13-17 %@ 2160-5629 %D 2014 %I Scientific Research Publishing %R 10.4236/oju.2014.42003 %X

Objectives: The Solyx System was developed to be easier and safer to use than other slings. It was the objective of this study to retrospectively assess the long-term safety and efficacy of the SolyxTM SIS Sling System. Methods: After IRB approval and informed consent, chart reviews with follow up phone questionnaires of 69 subjects implanted with the Solyx Sling were collected at 2 sites. All of the patients had SUI and had urethral hypermobility with a q-tip test of >30 degrees. All subjects underwent surgery from 12/2008 to 01/2010 with a mean follow up of 43 months (range 39 -49). Subjects included in this data collection had a mean age of 67 years (range 30 -87). The dominant type of incontinence within the study patients was SUI while 17/69 (25%) of the subjects also had a component of urge incontinence. 38/69 (55%) of study patients had concomitant procedures. Results: Long-term Solyx results showed 64/69 (93%) of patients were subjectively dry by questionnaire and were satisfied with their outcome. 63/69 (91%) would have the procedure again. There were 4 cases of denovo urge incontinence and 2 reports of transient retention. There were no serious adverse events including no bladder, bowel, vessel or nerve perforations and no erosions or extrusions. No pain was reported that was attributed to the implant. Conclusions: Chart review with follow phone questionnaires indicated that the Solyx Sling was a safe, efficacious and less-invasive option for patients requiring SUI surgery and that these results were sustainable for an average of 43 months.

%K Single Incision Sling %K Stress Urinary Incontinence %U http://www.scirp.org/journal/PaperInformation.aspx?PaperID=42878