Determination of Zoledronic Acid and Four Common Anions in Zoledronic Acid Injection by Gradient Ion Chromatography with a Self-Regeneration Suppressor
A simple method to determine zoledronic acid and
impurities in drug by ion chromatography with conductivity detection was
developed. Target analytes have been sufficiently separated from each other on
an Ionpac AS19 by using potassium hydroxide as eluent with a self-regeneration
suppressor. This novel proposed method enables direct analysis of zoledronic
acid in injection sample and provides good resolution. This novel method also exhibits
satisfactory precision (less than 1.68%) and accuracy (90.0% - 110.0%), good
linearity (r2 = 0.9993 - 0.9999) and sufficient
sensitivity (0.007 - 0.030 μg/mL). It can
be easily and conveniently adopted for the routine quality control analysis.
Cite this paper
Chen, M. and Ye, M. (2014). Determination of Zoledronic Acid and Four Common Anions in Zoledronic Acid Injection by Gradient Ion Chromatography with a Self-Regeneration Suppressor. Open Access Library Journal, 1, e827. doi: http://dx.doi.org/10.4236/oalib.1100827.
Major, P.P. and Cook, R. (2002) Efficacy of Bisphosphonates in the Management of
Skeletal Complications of Bone Metastases and Selection of Clinical Endpoints. American Journal of Clinical
Oncology, 25, 10. http://dx.doi.org/10.1097/00000421-200212001-00003
Xie, Z., Jiang, Y. and Zhang, D.Q. (2006) Simple Analysis of four
Bisphosphonates Simultaneously by Reverse Phase Liquid Chromatography Using
N-Amylamine as Volatile Ion-Pairing Agent. Journal of Chromatography A, 1104, 173. http://dx.doi.org/10.1016/j.chroma.2005.11.113
Lovdahl, M.J. and Pietrzyk, D.J. (1999)Anion-Exchange Separation and Determination of
Bisphosphonates and Related Analytes by Post-Column Indirect Fluorescence
Detection. Journal of Chromatography A, 850, 143. http://dx.doi.org/10.1016/S0021-9673(99)00622-6
Zhang, J.J. and Zhu, Y. (2007) Determination
of Betaine, Choline and Trimethylamine in Feed Additive by Ion-Exchange
Liquid Chromatography/Non-Suppressed Conductivity Detection. Journal of
Chromatography A, 1170, 114. http://dx.doi.org/10.1016/j.chroma.2007.09.014
Chen, M.L., Yan, W., Chen, Y.X., Jia, Z.S. and Zhu, Y. (2007) Determination of Impurities in Flame Retardant Monomer
2-Carboxyl Ethyl(Phenyl) Phosphinic Acid by Ion Chromatography. Journal of
Chromatography A, 1155, 47. http://dx.doi.org/10.1016/j.chroma.2007.02.045